| Model Number |
CQ2015A |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
03/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This product is not marketed in the us.Product evaluation found that the lens was returned in liquid in a lens vial.Visual inspection found haptic bent and piece of haptic missing.No similar complaint type events were reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon attempted to implant a cq2015a, +22.5 diopter, intraocular lens into the patient's eye on (b)(6) 2018.It was reported that the lens was loaded incorrectly.There was no patient contact.Attempts to obtain additional information have not been forthcoming.
|
| |
|