• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. PLATINUM 1 SERIES; SURGICAL ADJUNCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. PLATINUM 1 SERIES; SURGICAL ADJUNCTS Back to Search Results
Model Number 1MTEC30
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
This is not an implantable device.An attempt has been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that the doctor had some type of material coming out when inserting the lens.No patient injury reported.No further information provided.
 
Manufacturer Narrative
Device available for evaluation: yes.Returned to manufacturer on: 8/2/2018.Device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual evaluation and inspection using magnification was performed.Stress marks were observed at the cartridge tip.Traces of viscoelastic were also observed at the cartridge tube and neck section.The sample returned was not missing any part of the tip.The reported issue was not verified.There is no evidence to suggest that the complaint sample has been affected by the manufacturing process.No product deficiency was identified.The complaint cannot be confirmed.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no other complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATINUM 1 SERIES
Type of Device
SURGICAL ADJUNCTS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7457284
MDR Text Key106408845
Report Number2648035-2018-00600
Device Sequence Number1
Product Code KYB
UDI-Device Identifier05050474540323
UDI-Public(01)05050474540323(17)190118(10)CD00728
Combination Product (y/n)Y
PMA/PMN Number
K081545
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/01/2005,09/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date01/18/2019
Device Model Number1MTEC30
Device Catalogue Number1MTEC30
Device Lot NumberCD00728
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2018
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 03/27/2018
Initial Date FDA Received04/25/2018
Supplement Dates Manufacturer Received08/23/2018
Supplement Dates FDA Received09/20/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
LENS MODEL ZCB00 +23.0 DIOPTER SN (B)(4); LENS MODEL ZCB00 +23.0 DIOPTER SN (B)(4)
-
-