| Model Number |
LF4318 |
| Medical Device Problem Codes |
Overheating of Device (1437); Flare or Flash (2942)
|
| Health Effect - Clinical Code |
Partial thickness (Second Degree) Burn (2694)
|
| Date of Event |
04/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a procedure, the surgeon and staff both observed that the device was extremely hot to touch after activation.Accordingly, they said that the heat of the jaws caused two second degree burns (blisters) on the patient¿s right side of the incision with approximately 1cm in size each.The device was not being activated against the skin.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one device was received for evaluation.This device had been used in the treatment or diagnosis of a patient.A review of the lot number reported indicates that the product was within the assigned expiration date at the time of the reported incident.The returned product met specification as received.Visual inspection found no defects.The reported conditions were not confirmed.Functional testing was performed with the returned device on porcine kidney tissue.Multiple seals on various size vessels were made and a full thermal effect was visually verified.The device was activated while pressing the button and with the rotation knob in various positions to detect any problematic activation issues.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.No abnormal heat or burnt tissue was observed during the testing of the device.The alleged incident could not be duplicated.The investigation found the device to function normally and within specifications.The ifu (instructions for use) states, keep the instrument jaws clean.Build up of eschar may reduce sealing and/or cutting effectiveness.Wipe jaw surfaces and edges with a wet gauze pad as needed.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a procedure, the surgeon and staff both observed that the device was extremely hot to touch after activation.Accordingly, they said that the heat of the jaws caused two second degree burns (blisters) on the patient¿s right side of the incision with approximately 1cm in size each.The patient treated with a silvadene cream.The device was not being activated against the skin.
|
| |
|