| Model Number |
419488 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during implant the left ventricular (lv) lead dislodged due to patient anatomy.The lv lead was explanted and replaced.The patient is a participant in the post approval clinical surveillance product surveillance registry.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the full lead was returned, analyzed, and no anomalies were found.The outer insulation of the lead became extrinsically distorted due to kinking/buckling.Visual analysis of the lead indicated damage at implant.If information is provided in the future, a supplemental report will be issued.
|
| |
|