| Model Number |
D334DRG |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Delayed Charge Time (2586)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/21/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory indicated an alert for excessive charge time eos (end of service).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the implantable cardioverter defibrillator (icd) prematurely reached end of service (eos) due to excessive charge time.The device was explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was returned and analyzed.Analysis of the device found high internal battery resistance.Hybrid analysis confirmed that the device did not charge efficiently by observing an excessive charge time as-received.The device charged nominally when the battery was replaced with an external supply.No hybrid anomalies were found.Battery analysis revealed that the reported excessive charge times likely resulted from a spurious increased battery resistance induced by lithium-severing, a known failure mode for 35 joule batteries rejected from the device was received and analyzed.If information is provided in the future, a supplemental report will be issued.
|
| |
|