• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD Back to Search Results
Catalog Number 4004C1440
Medical Device Problem Codes Kinked (1339); Device Handling Problem (3265)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/29/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2018-00851, 3005168196-2018-00852.The hospital disposed of the device.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure for a vessel sacrifice using penumbra coils 400 (pc400¿s) and a penumbra coil 400 detachment handle (handle).During the procedure, while advancing a pc400 through a px slim delivery microcatheter (px slim), the hospital fellow was too aggressive and inadvertently bent the pc400 pusher assembly; therefore, it was removed.It was reported that another pc400 pusher assembly was also bent while advancing the pc400 through the px slim; therefore, it was removed.The procedure continued using additional coils.Around the 30th coil, it was reported that blood might have gotten inside the handle and its the trigger became a bit sticky; therefore, the handle was set aside.The procedure was completed using the same px slim, a new handle and additional pc400's.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA COIL 400
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7461671
Report Number3005168196-2018-00850
Device Sequence Number19145550
Product Code HCG
UDI-Device Identifier00814548011059
UDI-Public00814548011059
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date07/01/2020
Device Catalogue Number4004C1440
Device Lot NumberA28772
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2018
Initial Report FDA Received Date04/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age39 YR
-
-