Model Number DTBA1QQ
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during the implant procedure, when the physician went to place the atrial lead, the setscrew would not engage the lead and kept turning without clamping down on the lead.The physician attempted this three or four more times pulling the wrench out and re-engaging the grommet; but it did not work.Another wrench was tried, but the setscrew still would not ratchet on to the atrial lead.When the wrench was pulled out, the setscrew came out also.At that point a new device was attempted and implanted with no issues.Also, when pulling the right ventricular lead out of the first device, the setscrew came out with the wrench as well.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.The returned device had a damaged connector.The returned device indicated a grommet issue.The returned device indicated a damaged grommet.The returned device indicated a missing set screw.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA QUAD XT
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH  1131
MDR Report Key7461894
Report Number9614453-2018-01658
Device Sequence Number11036122
Product Code NIK
UDI-Device Identifier00763000059651
UDI-Public00763000059651
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P010031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/28/2019
Device Model NumberDTBA1QQ
Device Catalogue NumberDTBA1QQ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2018
Initial Date Received by Manufacturer 03/22/2018
Supplement Date Received by Manufacturer05/14/2018
Initial Report FDA Received Date04/25/2018
Supplement Report FDA Received Date05/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age55 YR
-
-