Product analysis: analysis could not confirm the reported issue; output testing at multiple settings noted no anomalies.The external pulse generator (epg) failed incoming testing for atrial pace pulse noise.It was found that the upper and lower cases, display flex, display frame, keypad flex, main printed circuit board (pcb) and six main pcb screws were all contaminated.It was also noted that the display wires were pinched, but the insulation was not compromised.All found defective parts were replaced and all other identified issues were resolved.The epg passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
|