| Model Number |
5554-53 |
| Medical Device Problem Codes |
Decreased Sensitivity (2534); Device Dislodged or Dislocated (2923); High Capture Threshold (3266)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/28/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the right atrial (ra) lead exhibited low p-waves and high thresholds.It was confirmed by x-ray that the lead had dislodged.The lead was explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the full lead was returned, analyzed, and no anomalies were found.If information is provided in the future, a supplemental report will be issued.
|
| |
|