• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA STEM COLLARED HA COATED STD STEM SIZE 3; CEMENTLESS HIP STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA STEM COLLARED HA COATED STD STEM SIZE 3; CEMENTLESS HIP STEM Back to Search Results
Catalog Number 01.18.233
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 05/29/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 25 april 2018.Lot 170844: (b)(4) items manufactured and released on (b)(4) 2017.Expiration date: 2022-03-28 no anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Others devices involved: versafitcup 01.26.48mb (k083116) acetabular shell ø 48; lot 166338: (b)(4) items manufactured and released on (b)(4) 2017.Expiration date: 2022-01-10.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Versafitcup dm 01.26.2848mhc (k092265) double mobility hc liner ø 48/28; lot 166698: (b)(4) items manufactured and released on (b)(4) 2017.Expiration date: 2022-01-01.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold with two similar reported event.Mectacer 01.29.201 (k112115), biolox delta ceramic ball head 12/14 ø 28 size s -3.5; lot 170679: (b)(4) items manufactured and released on (b)(4) 2017.Expiration date: 2022-05-14 no anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
 
Event or Problem Description
The patient came in due to signs of infection 7 months after primary.The patient was (b)(6) for (b)(6).All implants were removed and a spacer was implanted.The surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEM COLLARED HA COATED STD STEM SIZE 3
Common Device Name
CEMENTLESS HIP STEM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
MDR Report Key7464053
Report Number3005180920-2018-00278
Device Sequence Number1630181
Product Code LZO
UDI-Device Identifier07630030804588
UDI-Public07630030804588
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K121011
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/28/2022
Device Catalogue Number01.18.233
Device Lot Number170844
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2018
Initial Report FDA Received Date04/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-