MAUDE Adverse Event Report: BAXTER HEALTHCARE - DOMINICAN REPUBLIC CLEARLINK BURETROL SOLUTION SETS; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - DOMINICAN REPUBLIC CLEARLINK BURETROL SOLUTION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2H8864
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that air entered the tubing of two (2) clearlink system buretrol solution sets.At the time of the reported problem, a lactated ringer solution bag was being drawn through the set; however, the process step and patient involvement was not specified.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
The actual device was not available; however, eight (8) photographs of the sample were provided for evaluation.The photographs were inspected and air bubbles were visible in the tubing below the drip chamber.The reported condition was verified.The cause of the reported condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK BURETROL SOLUTION SETS
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
haina, san cristobal
MDR Report Key7464592
Report Number1416980-2018-02390
Device Sequence Number11148987
Product Code FPA
UDI-Device Identifier00085412071749
UDI-Public(01)00085412071749
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K123874
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date12/22/2022
Device Catalogue Number2H8864
Device Lot NumberDR17L20047
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/02/2018
Supplement Date Received by Manufacturer05/14/2018
Initial Report FDA Received Date04/26/2018
Supplement Report FDA Received Date05/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LACTATED RINGER SOLUTION BAG, 1000ML
-
-