| Model Number |
8886621741 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned products noted one suture and one needle.The needle was observed to be detached from the suture.Upon microscopic inspection of the needle, normal needle crimp and swage marks were observed on the needle.The retainer was inspected with no abnormalities.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intra-operatively, the device needle was detached form the thread.The status of the patient was fine.
|
| |
|
Search Alerts/Recalls
|