Model Number 010911L
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Codes Tissue Damage (2104); No Known Impact Or Consequence To Patient (2692)
Date of Event 04/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during laparoscopic donor nephrectomy (left side), the device was able to fire, but the staples on the reload and tissue are malformed.The doctor fired the device on the renal vein (3rd firing).On first firing, ureter, second, renal artery, third: rental vein.Upon completion of firing, the doctor attempted to open the stapler but the tissue/staples were stuck to the reload.The doctor was able to fire two clips below the jaws of the reload and had to cut the tissue to remove the tissue from the stapler.There was unanticipated tissue loss and tissue damage a result of the issue.A loss of 1-2mm tissue was noted.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of a photo of a device.Staple formation issues were observed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTIFIRE ENDO TA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key7465169
Report Number2647580-2018-02118
Device Sequence Number11161017
Product Code GDW
UDI-Device Identifier20884521047768
UDI-Public20884521047768
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K910192
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2021
Device Model Number010911L
Device Catalogue Number010911L
Device Lot NumberP6J0362MX
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/17/2018
Supplement Date Received by Manufacturer07/13/2018
08/08/2018
Initial Report FDA Received Date04/26/2018
Supplement Report FDA Received Date07/30/2018
10/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1