| Model Number |
H7493926024350 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
03/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product.(b)(4).Device evaluated by mfr: stent delivery system was returned for analysis.A visual examination of the stent found that the first proximal stent row was damaged with stent struts lifted and pulled distally.The undamaged crimped stent outer diameter was measured and was within specification.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple hypotube kinks along the full length of the catheter.This type of damage is consistent with excessive force being applied to the delivery system.A visual and tactile examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is caused by other device as another device/drug/subsequent procedure caused the complaint event.(b)(4).
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 05-april-2018.It was reported that device interaction occurred.During introduction of a 3.50 x 24mm synergy ii drug-eluting stent, the stent strut got caught in the hemostatic valve.The stent was removed and it never went inside the patient's body.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was well.However, returned device analysis revealed proximal stent damaged.
|
| |
|
Search Alerts/Recalls
|