• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ACCUTRON DIGITAL ULTRA FLOWMETER; NITROUS MIXER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ACCUTRON DIGITAL ULTRA FLOWMETER; NITROUS MIXER Back to Search Results
Model Number 56100
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code Partial thickness (Second Degree) Burn (2694)
Date of Event 03/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A facility reported that the accutron nitrous mixer unit caught fire and injured a dental assistant employee while switching from the first oxygen tank to the second oxygen tank.This employee suffered 2nd degree burns to both the left hand and right thigh.That same day, they were sent to burn unit at uab where they received wound care treatment.The employee was discharged the same day and given instructions on wound care treatment to perform.Accutron qa followed up directly with the facility.It was confirmed that there was no patient in the room and no other personnel was injured as a result of this incident.Dr.(b)(6), the facility dentist who was in the next room indicated that the fire came out of the front near the "empty" cylinder upon changing over to the reserve back cylinder.The fire depleted quickly and no external extinguishing methods were necessary.Dr.(b)(6) also reported that he identified a gas leak with the equipment a week earlier before the incident.There was no reported safety actions taken to resolve or investigate the source of the gas leak before the incident occurred which is not compliant with the nitrous mixer's ifu required training on nitrous oxide use.After detailed review of the provided information and condition of the equipment returned, there is potential that the check valve became contaminated (particulate debris) during their equipment transport between various offices and did not fully close.The reserve cylinder having a (contaminated) check valve could cause a gas leak.Dr.(b)(6) indicated that the cylinders are taken off the manifold during routine transport between facilities which could potentially expose the manifold ports to contaminants.Dr.(b)(6) stated that he does not think the problem was with the nitrous mixer unit.The oxygen cylinder is currently under investigation by (b)(4), distributor/filler of the gas cylinders.This is an isolated incident that will continue to be monitored in accutron's complaint handling system.(b)(6).
 
Event or Problem Description
A facility reported that the accutron nitrous mixer unit caught fire and injured a dental assistant employee while switching from first oxygen tank to the second oxygen tank.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIGITAL ULTRA FLOWMETER
Common Device Name
NITROUS MIXER
Manufacturer (Section D)
ACCUTRON
1733 west parkside ln
pheonix AZ 85027
MDR Report Key7465840
Report Number2020813-2018-00001
Device Sequence Number10803109
Product Code BSZ
UDI-Device Identifier00813830023374
UDI-Public00813830023374
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K052335
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Other
Remedial Action Replace,Patient Monitoring
Type of Report Initial
Report Date (Section B) 04/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Dental Assistant
Device Model Number56100
Device Lot Number9286
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/28/2018
Initial Report FDA Received Date04/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AIRGAS OXYGEN FILLED CYLINDER
Outcome Attributed to Adverse Event Other;
Patient Age26 YR
-
-