| Catalog Number |
100/897/090CZ |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One portex ® uniper® adjustable flange tracheostomy tube was returned for analysis in used condition.Visual inspection of the tracheostomy tube showed no issues.The inflation test was repeated on the received sample and no leakage was observed after 12 hours.
|
| |
|
Event or Problem Description
|
|
It was reported that "balloon lost tautness slowly after inflation" after intubation.No serious injury was reported in connection with this incident.
|
| |
|
Search Alerts/Recalls
|