| Catalog Number |
306521 |
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
04/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.Multiple potential lot numbers were provided for this incident.The information for each lot number is as follows: medical device lot #: 719291n.Medical device expiration date: 1/10/2019.Device manufacture date: 7/17/2017.Medical device lot #: 726971n.Medical device expiration date: 3/25/2019.Device manufacture date: 9/29/2017.Medical device lot #: 726391n.Medical device expiration date: 3/19/2019.Device manufacture date: 9/26/2017.Medical device lot #: 726472n.Medical device expiration date: 3/20/2019.Device manufacture date: 9/26/2017.Medical device lot #: 718091n.Medical device expiration date: 12/2/2018.Device manufacture date: 7/8/2017.Medical device lot #: 721481n.Medical device expiration date: 2/1/2019.Device manufacture date: 8/8/2017.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that infections associated with central lines and the use of a 3 ml bd posiflush¿ heparin lock flush syringe, in 10 ml syringe, 10 usp units/ml were being investigated by the fda.There was no evidence that the bd device is implicated but all sources and common products are being investigated.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: type of report: the initial mdr was submitted as a 30-day report.Bd has changed our position on this incident and we are making a correction to report the mdr as a 5-day report.
|
| |
|
Event or Problem Description
|
|
It was reported that infections associated with central lines and the use of a 3 ml bd posiflush¿ heparin lock flush syringe, in 10 ml syringe, 10 usp units/ml were being investigated by the fda.There was no evidence that the bd device is implicated but all sources and common products are being investigated.
|
| |
|
Additional Manufacturer Narrative
|
|
Lot numbers 718091n, 719291n, 721481n, 726391n, 726472n, and 726971n for product code 306521 were provided for evaluation by our quality engineer team.Upon reviewing the production history for the provided lot numbers, no deviations or non-conformances were identified during the manufacturing process.A corrective and preventive action plan was initiated to further investigate and monitor this issue.One-hundred and twenty retained samples for each lot number provided were visually inspected and no abnormalities were observed in regards to the solution.Forty retained samples for each lot number provided were sent for sterility testing.The sterility testing confirmed that the samples exhibited no microbial growth after incubation.A direct causation between the reported infections and the bd franklin product has not been identified.Prior to the report of infection evaluated in (b)(4), there was no trend for infection cases reported for franklin product.The complaint is part of a new trend which began in april 2018.Capa (b)(4) was initiated to address this issue.
|
| |
|