MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS 5ML BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGE, IN 10ML SYRINGE, 100 USP UNITS/ML; PREFILLED HEPARIN FLUSH SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS 5ML BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGE, IN 10ML SYRINGE, 100 USP UNITS/ML; PREFILLED HEPARIN FLUSH SYRINGE Back to Search Results
Catalog Number 306513
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 04/10/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The date received by manufacturer has been used for this field.Multiple potential lot numbers were provided for this incident.The information for each lot number is as follows: medical device lot #: 716192n medical device expiration date: 6/9/2019 device manufacture date: 6/15/2017 medical device lot #: 718711n medical device expiration date: 7/5/2019 device manufacture date: 7/8/2017 medical device lot #: 718611n medical device expiration date: 7/4/2019 device manufacture date: 7/7/2017 medical device lot #: 725181n medical device expiration date: 9/7/2019 device manufacture date: 9/12/2017 medical device lot #: 725271n medical device expiration date: 9/8/2019 device manufacture date: 9//12/2017 a device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that infections associated with central lines and the use of a 5 ml bd posiflush¿ heparin lock flush syringe, in 10 ml syringe, 100 usp units/ml were being investigated by the fda.There was no evidence that the bd device is implicated but all sources and common products are being investigated.
 
Additional Manufacturer Narrative
Correction: type of report: the initial mdr was submitted as a 30-day report.Bd has changed our position on this incident and we are making a correction to report the mdr as a 5-day report.
 
Event or Problem Description
It was reported that infections associated with central lines and the use of a 5 ml bd posiflush¿ heparin lock flush syringe, in 10 ml syringe, 100 usp units/ml were being investigated by the fda.There was no evidence that the bd device is implicated but all sources and common products are being investigated.
 
Additional Manufacturer Narrative
H.6.Investigation summary: lot: 725181n & 725271n & 716192n & 718711n & 718611n.There were no deviations, non-conformances, or out of specifications related to the manufacture of lot 725181n or the raw materials used to manufacture this lot.This complaint is part of a new trend which began in (b)(6) 2018.Capa (b)(4) was initiated to address this issue.100% of retained samples for the lot ((b)(4) units) were visually inspected.No growth was seen (i.E., solution was clear).(b)(4) units of the retained samples were sent to nelson laboratories, inc.For sterility testing in may 2018.The sterility testing was performed in an isolator per usp 71 membrane filtration method.Per the attached lab report, after incubation, the retained samples exhibited no microbial growth.Investigation conclusion: the root cause analysis of the reported infection cases under capa (b)(4) has not identified a direct causation between the infections and the bd franklin product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5ML BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGE, IN 10ML SYRINGE, 100 USP UNITS/ML
Common Device Name
PREFILLED HEPARIN FLUSH SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
9630 south 54th street
franklin WI 53132
MDR Report Key7467625
Report Number2134319-2018-00008
Device Sequence Number1273449
Product Code FOZ
UDI-Device Identifier30382903065135
UDI-Public30382903065135
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K011967
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Catalogue Number306513
Device Lot NumberSEE H.10.
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/10/2018
Supplement Date Received by Manufacturer04/10/2018
04/10/2018
Initial Report FDA Received Date04/27/2018
Supplement Report FDA Received Date05/01/2018
08/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-