| Model Number |
MMT-551PNAB |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump alarmed compromised force sensor system.The customer¿s blood glucose level was 116 mg/dl at the time of the incident.The customer stated that the top of the reservoir was cracked.The customer stated that the pump squirted out insulin.The customer was advised to discontinue use of the pump and was advised to revert to back up plan per health care professional instructions.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump passed the displacement test but compromised force sensor system alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform the occlusion, prime/compromised force sensor system and excessive no delivery test due to compromised force sensor system alarm.
|
| |
|