• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MALLINCKRODT MEDICAL SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MALLINCKRODT MEDICAL SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 70XLTCD
Medical Device Problem Code Air Leak (1008)
Health Effect - Clinical Code Extubate (2402)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, two trach tubes were leaking and unable to maintain air in the balloon.The customer reported that there was no known patient involvement and outcome.
 
Additional Manufacturer Narrative
Evaluation summary: one sample returned for evaluation.Visual examination shows no sign of any defect.The returned unit was inflated and leak tested under water.No leakage detected.The unit remained inflated for a four-hour period and no leakage was detected.The reported defect could not be detected on the unit returned for evaluation.The root cause of the reported defect could not be determined as no defect was detected on the unit returned for evaluation.All of our cuffed extended-length tracheostomy tubes(xlt) product undergo a four hour inflate/deflate test and it is at this stage of the process that any defect is removed from the lot.The product relates to the reported complaint event.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MALLINCKRODT MEDICAL
cornamaddy
athlone 3810
MDR Report Key7468612
Report Number8020889-2018-00048
Device Sequence Number13761396
Product Code JOH
UDI-Device Identifier20884522028162
UDI-Public20884522028162
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K051416
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/01/2021
Device Model Number70XLTCD
Device Catalogue Number70XLTCD
Device Lot Number201610052X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2018
Initial Date Received by Manufacturer 04/19/2018
Supplement Date Received by Manufacturer09/04/2018
Initial Report FDA Received Date04/27/2018
Supplement Report FDA Received Date10/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-