ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN@HOME, RF TELEMETRY BASIC PACEMAKER DATA TRANSMITTER Back to Search Results
Model Number EX1150
Device Problems Thermal Decomposition of Device (1071); Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/23/2018
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis. Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the merlin @ home transmitter was unable to power up. Upon investigation, the transmitter's prongs exhibited charring. The cause of the issue was determined to be damage from a lightning strike. The transmitter was replaced.
 
Manufacturer Narrative
Final analysis found charring on the prongs of the ac power adapter, burnt components on the main printed circuit board of the adapter, and a burnt mark on the inner casing of the ac power adapter. Burnt components were also noted on the main printed circuit board of the transmitter and a burn mark on the inner housing of the transmitter, all of which contributed to the field complaint of power up anomaly. The transmitter also had a burnt mark on its inner housing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMERLIN@HOME, RF TELEMETRY BASIC
Type of DevicePACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7468745
MDR Text Key106747950
Report Number2017865-2018-06450
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 06/28/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberEX1150
Device Catalogue NumberEX1150
Other Device ID Number05414734504799
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2018
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received06/27/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-