| Model Number |
VSR100005 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during laparoscopic excision of lesion of uterus, it was not too smooth when inserting the dilator into the outer sheath of the device.The surgical time was extended by less than 30 min.The procedure was completed with another device.The status of the patient was reported as no problem.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted the protective sheath of the radially expandable sleeve was cut.The trocar and envelope seal appeared intact.The condition in which the device was received precludes functional testing.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of the cut in the radially expandable sleeve may occur when mishandled during clinical application.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.If information is provided in the future, a supplemental report will be issued.
|
| |
|