• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD VERSASTEP; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD VERSASTEP; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number VSR100005
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during laparoscopic excision of lesion of uterus, it was not too smooth when inserting the dilator into the outer sheath of the device.The surgical time was extended by less than 30 min.The procedure was completed with another device.The status of the patient was reported as no problem.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted the protective sheath of the radially expandable sleeve was cut.The trocar and envelope seal appeared intact.The condition in which the device was received precludes functional testing.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of the cut in the radially expandable sleeve may occur when mishandled during clinical application.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSASTEP
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7469387
Report Number9612501-2018-00877
Device Sequence Number18863970
Product Code GCJ
UDI-Device Identifier10884521084209
UDI-Public10884521084209
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K012539
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model NumberVSR100005
Device Catalogue NumberVSR100005
Device Lot NumberJ7K1067X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2018
Initial Date Received by Manufacturer 04/18/2018
Supplement Date Received by Manufacturer06/20/2018
Initial Report FDA Received Date04/27/2018
Supplement Report FDA Received Date07/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-