Model Number N/A |
Device Problems
Device Contamination with Chemical or Other Material (2944); Packaging Problem (3007)
|
Patient Problem
No Information (3190)
|
Event Date 04/02/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).(b)(6).Reporter had indicated that product will not be returned.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that foreign substances were inside of the sterile packages.No adverse events have been reported as a result of the malfunction.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned products identified that there is debris inside the sterile package.Review of the device history records identified no related deviations or anomalies.These products likely left zimmer biomet control non-conforming.The root cause of the reported issue is attributed to manufacturing deficiency.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|