• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number MICL13.2
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 02/15/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Secondary surgical intervention, repositioning, explant and exchange for a shorter lens.Claim # (b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon implanted a 13.2 mm micl13.2, -10.5 diopter, implantable collamer lens into the patient's left eye (os) on (b)(6) 2018.The lens was removed on (b)(6) 2018 due to observation of excessive vault with glare/haloes and blurred vision.It was reported that the lens was repositioned but it did not resolve the problem.On (b)(6) 2018 the lens was exchanged for a shorter length lens and the problem was resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7470285
Report Number2023826-2018-00623
Device Sequence Number114035
Product Code MTA
UDI-Device Identifier00841542103305
UDI-Public00841542103305
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date07/31/2019
Device Model NumberMICL13.2
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2018
Initial Report FDA Received Date04/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age39 YR
-
-