| Model Number |
MICL13.2 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
02/15/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Secondary surgical intervention, repositioning, explant and exchange for a shorter lens.Claim # (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 13.2 mm micl13.2, -10.5 diopter, implantable collamer lens into the patient's left eye (os) on (b)(6) 2018.The lens was removed on (b)(6) 2018 due to observation of excessive vault with glare/haloes and blurred vision.It was reported that the lens was repositioned but it did not resolve the problem.On (b)(6) 2018 the lens was exchanged for a shorter length lens and the problem was resolved.
|
| |
|