Clinical evaluations was able to find substantial evidence to support the following reported events; intraoperative endoleak difficult to deploy.At implant, these events were confirmed by medical documentation only: difficult to deploy, an unknown endoleak (no imaging to confirm the type iiib endoleak), the persistent type ii endoleak and, significant blood loss that required transfusions.The difficulty removing the outer sleeve, and the unusual amount of tension needed to deploy the stent might have contributed to this complaint.The components underwent many balloon angioplasties due to the transient type ib endoleak, and some areas with a 32 mm coda balloon, which might also have contributed to this complaint.The patient was discharged home on pod1.A root cause investigation was carried out for all afx complaints having an identified failure mode of a type iiib endoleak.Endologix implemented the following corrective actions with the intent of reducing type iiib endoleak events; 1.Upgraded graft material (i.E.(b)(4)) and 2.Updates to the ifu and additional physician training.The change to (b)(4) graft material and the ifu changes were put in place july 2014.The type iiib endoleak rate for afx manufactured and implanted before these corrective actions were put in place is trending at 2.5%.Since the corrective actions were implemented, the type iiib endoleak events reported for afx devices has been reduced to <0.2%.Root cause investigation was carried out for all afx complaints having an identified failure mode of a type 3b endoleak.Endologix implemented the following corrective actions with the intent of reducing type 3b endoleak events; 1.Upgraded graft material (i.E.(b)(4) ) and 2.Updates to the ifu and additional physician training.The change to (b)(4) graft material and the ifu changes were put in place july 2014.The type 3b endoleak rate for afx manufactured and implanted before these corrective actions were put in place is trending at 2.5%.Since the corrective actions were implemented, the type 3b endoleak events reported for afx devices has been reduced to <0.2%.Devices remain implanted in the patient and were not returned, no evaluation completed.The review of manufacturing lot confirmed all devices met specifications prior to release.Endologix continues to investigate this event and similar events to ensure the highest quality and patient safety.
|