• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S TITAN TOUCH PUMP; INFLATABLE PENILE PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S TITAN TOUCH PUMP; INFLATABLE PENILE PROSTHESIS Back to Search Results
Model Number 5177502400
Device Problem Torn Material (3024)
Patient Problem No Information (3190)
Event Date 03/30/2018
Event Type  Injury  
Manufacturer Narrative
Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Once our evaluation is complete, a follow-up report will be submitted.
 
Event Description
According to the available information, titan was explanted due to a tubing break at pump junction.
 
Manufacturer Narrative
This follow up mdr is created to document the conclusion of the investigation.Requests for the explanted prosthesis and for additional information surrounding this event have been made.However, to date neither the device nor the information has been received.Without the benefit of analyzing the explant and without the requested information, quality cannot comment on the condition of the device or the events surrounding this incident.Quality reviewed the manufacturing records for this device and confirmed that there were no discrepancies that would have contributed to this complaint and verified that the devices from this lot met all specifications prior to release.Review of nonconforming reports revealed no nonconformance with this lot that would have contributed to the event.A review of the complaint database revealed no significant trends in complaints of this type for lot 5067749.The most likely root cause of the event cannot be determined at this time.If the explanted device or additional information is received, quality will re-evaluate this event in accordance with procedures.Management routinely reviews events such as this and monitors complaint levels.Additionally, events of this type are captured in the product risk documentation.Based on this information, no further corrective action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAN TOUCH PUMP
Type of Device
INFLATABLE PENILE PROSTHESIS
Manufacturer (Section D)
COLOPLAST A/S
holtedam 1
humlebaek, 3050
DA  3050
Manufacturer (Section G)
COLOPLAST MANUFACTURING US, LLC
1601 west rivier road north
minneapolis MN 55411
Manufacturer Contact
stephanie perryman
1601 west rivier road north
minneapolis, MN 55411
6124345685
MDR Report Key7470571
MDR Text Key106831786
Report Number2125050-2018-00321
Device Sequence Number1
Product Code FHW
UDI-Device Identifier05708932533966
UDI-Public05708932533966
Combination Product (y/n)N
PMA/PMN Number
P000006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 06/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number5177502400
Device Catalogue Number5177502400
Device Lot Number5067749
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/30/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age41 YR
-
-