• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 40 PARKER SEDLINE EEG SENSOR; MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 40 PARKER SEDLINE EEG SENSOR; MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS) Back to Search Results
Model Number 4248-9
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Event or Problem Description
The customer reported: "psi value shows over 80 even though patient is well anesthetized".No consequences or impact to patient was reported.
 
Additional Manufacturer Narrative
The returned sensor was evaluated.During evaluation, the sensor passed visual inspection; however, failed continuity testing due to open cb, ct, l2, l1, r1, r2 across the electrodes.Sensor was returned used and could not be tested on forehead for functional testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEDLINE EEG SENSOR
Common Device Name
MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS)
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
MDR Report Key7471278
Report Number2031172-2018-00250
Device Sequence Number144971
Product Code MWI
UDI-Device Identifier10843997005804
UDI-Public10843997005804
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K151644
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/01/2019
Device Model Number4248-9
Device Catalogue Number2479
Device Lot Number17K14
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2018
Was the Report Sent to FDA? Yes
Initial Date Received by Manufacturer 04/04/2018
Supplement Date Received by Manufacturer05/11/2018
Initial Report FDA Received Date04/28/2018
Supplement Report FDA Received Date06/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-