| Model Number |
4248-9 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported: "psi value shows over 80 even though patient is well anesthetized".No consequences or impact to patient was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned sensor was evaluated.During evaluation, the sensor passed visual inspection; however, failed continuity testing due to open cb, ct, l2, l1, r1, r2 across the electrodes.Sensor was returned used and could not be tested on forehead for functional testing.
|
| |
|
Search Alerts/Recalls
|