| Model Number |
G52011 |
| Medical Device Problem Codes |
Kinked (1339); Material Perforation (2205)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
510(k) number: k160229.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
The needle and sheath "separated" after first "attempt" which affected second biopsy.Operator changed another same product to finish the biopsy.
|
| |
|
Event or Problem Description
|
|
This follow up report is being submitted to cancel the initial report.The needle and sheath seperated after first attemp which affected second biopsy.Operator changed another same product to finish the biopsy.
|
| |
|
Additional Manufacturer Narrative
|
|
510(k) number: k160229.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining (b)(4).Importer site establishment registration number: (b)(4).This follow up report is being submitted to cancel the initial mdr report.A conservative assessment deemed this event reportable based on the malfunction precedence "distal tip needle/ sheath perforation".However, following the device evaluation no perforation was observed.This event has been re-assessed and this report is to notify the fda that this event no longer meets the fda reporting criteria of a malfunction report.No adverse effects to the patient was reported as occurring.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|
Search Alerts/Recalls
|