Model Number TRUEMETRIX
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(manufacturer narrative = t, corrected data = f) (b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
 
Event or Problem Description
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results of 145, 166, 239 141 and 189 mg/dl.The customer's expected fasting blood glucose test result is 120 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was performed by the customer non-fasting and produced test results of 143 mg/dl and 154 mg/dl using truemetrix meter.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 11/30/2018 and open vial date is less than one month.The meter memory was reviewed for previous test result history (date/time not set): (b)(6).
 
Event or Problem Description
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results of 145, 166, 239 141 and 189 mg/dl.The customer's expected fasting blood glucose test result is 120 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was performed by the customer non-fasting and produced test results of 143 mg/dl and 154 mg/dl using truemetrix meter.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is (b)(6) 2018 and open vial date is less than one month.The meter memory was reviewed for previous test result history (date/time not set): result 1:145mg/dl date:(b)(6) 2018 time:6:39 am fasting.Result 2:166mg/dl date:(b)(6) 2017 time:11:31 am fasting.Result 3:239mg/dl date:(b)(6) 2017 time:12:04 pm fasting.Result 4:141mg/dl date:(b)(6) 2017 time:9:43 pm fasting.Result 5:189mg/dl date:(b)(6) 2017 time:2:23 pm fasting.
 
Additional Manufacturer Narrative
(manufacturer narrative = t, corrected data = f) internal report # 01556147 returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7472184
Report Number1000113657-2018-00494
Device Sequence Number16171502
Product Code NBW
UDI-Device Identifier00050428275283
UDI-Public(01)00050428275283
Combination Product (Y/N)N
Initial Reporter StateNJ
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 10/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/30/2018
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMU2491
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2018
Was the Report Sent to FDA? No
Event Location Home
Initial Date Received by Manufacturer 04/04/2018
Supplement Date Received by Manufacturer04/04/2018
Initial Report FDA Received Date04/30/2018
Supplement Report FDA Received Date10/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SECOND THERAPY; SECOND THERAPY
-
-