ETHICON INC. GUT CHR UD 27IN 3-0 S/A SH; SUTURE, ABSORBABLE, NATURAL Back to Search Results
Catalog Number G122H
Medical Device Problem Code Detachment Of Device Component (1104)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that a patient underwent a gynecological procedure in (b)(6) 2018 and suture was used.During the procedure, the needle prematurely detached from the suture.There were no adverse patient consequences.No additional information was provided.
 
Additional Manufacturer Narrative
Representative samples were returned for analysis.There were no defects found on the packages.The samples were opened and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strands and no defects were observed.Functional test was performed and the pull force were above the minimum requirements.Per the sample condition no needle pull off was found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUT CHR UD 27IN 3-0 S/A SH
Common Device Name
SUTURE, ABSORBABLE, NATURAL
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
MDR Report Key7472360
Report Number2210968-2018-72440
Device Sequence Number16169826
Product Code GAL
UDI-Device Identifier10705031037441
UDI-Public10705031037441
Combination Product (Y/N)N
Initial Reporter StateCA
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 04/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2022
Device Catalogue NumberG122H
Device Lot NumberLHM416
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2018
Initial Date Received by Manufacturer 04/04/2018
Supplement Date Received by Manufacturer04/19/2018
Initial Report FDA Received Date04/30/2018
Supplement Report FDA Received Date05/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-