• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION PROPAQ MD DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION PROPAQ MD DEFIBRILLATOR Back to Search Results
Model Number PROPAQ MD
Medical Device Problem Code Failure of Device to Self-Test (2937)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event or Problem Description
Complainant alleged that during biomed testing, the device failed self test for defib function.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
The complainant was contacted for return of the device.The device has not been returned to zoll for evaluation.
 
Additional Manufacturer Narrative
The reported malfunction was observed and attributed to a faulty transformer on the processor/bridge/pace board.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROPAQ MD DEFIBRILLATOR
Common Device Name
PROPAQ MD
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key7472872
Report Number1220908-2018-01133
Device Sequence Number157770
Product Code MKJ
UDI-Device Identifier00847946017682
UDI-Public00847946017682
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K100654
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberPROPAQ MD
Device Catalogue NumberPROPAQ MD
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2018
Initial Date Received by Manufacturer 04/09/2018
Supplement Date Received by Manufacturer04/09/2018
04/09/2018
Initial Report FDA Received Date04/30/2018
Supplement Report FDA Received Date05/23/2018
06/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-