| Model Number |
MT22608-PNK |
| Medical Device Problem Code |
Device Inoperable (1663)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(4) 2018 that on (b)(6) 2018 the patient experienced a receiver that ceased to function.The complaint receiver was returned for evaluation.The device was visually inspected and no defect was found.The receiver will boot and charge.The receiver download was retrieved and confirmed the complaint.Receiver functional tester: passed all relevant testing.The reported event of the device ceased to function was confirmed via log review.The probable cause is determined to be a defective battery.
|
| |
|