| Catalog Number |
B1070-020 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The product was not returned to abbott vascular for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.The investigation was unable to determine a conclusive cause for the reported difficulty to position the guide wire/sheath and difficulty removing the guide wire.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a non-tortuous and mildly calcified lesion located in the renal artery.A 7.0 x 20 mm armada 35 balloon catheter could not advance over an unspecified guide wire as resistance was felt through a 6fr introducer sheath.Resistance was met during the removal of the balloon catheter from the guide wire.The guide wire remained in the patient and a 4fr balloon catheter was successfully advanced over the guide wire instead to complete the procedure.There was no adverse patient sequela and no reported clinically significant delay in the procedure or therapy.No additional information was provided.
|
| |
|