• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION
Device Problem Device Stops Intermittently (1599)
Patient Problem No Patient Involvement (2645)
Event Date 04/05/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Device udi not provided.Device manufacturing date is unavailable.A medtronic representative went to the site to test the equipment.The manufacturing representative noted that the power cable was loose and tightened the cable and the attachment.The imaging system then passed the system checkout and was found to be fully functional.No parts have been received by the manufacturer for evaluation.
 
Event Description
Medtronic received information regarding a navigation device being used outside of a procedure.It was reported that the system lost power while importing an exam via cd, and it would not turn on afterwards.There was no patient present when this issue was identified.No further information was provided.
 
Manufacturer Narrative
Additional information: unique device identification (udi) and device manufacture date provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7473133
MDR Text Key106959890
Report Number1723170-2018-01801
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 06/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberFUSION
Device Catalogue Number9733560
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/05/2018
Initial Date FDA Received04/30/2018
Supplement Dates Manufacturer Received05/15/2018
Supplement Dates FDA Received06/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/07/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-