| Model Number |
M0035474060 |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During the coil embolization procedure, it was reported that the physician thought the first coil (subject device) was smaller than the size on the label.However, he still detached the coil.Then the physician advanced the second coil and big resistance was felt in the microcatheter.So he withdrew the second coil, but the first coil came out with the second coil together and the first coil was found stretched.The procedure was completed with other coils.No clinical consequences reported to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.During the analysis of the returned device, it was revealed that the main coil was kinked, stretched and the main coil suture was damaged.So it was impossible to measure the length of the coil; therefore the cause of the reported main coil incorrect dimension could not be determined.In addition, the delivery wire was revealed kinked due to handling.No functional test could be performed on the returned device.Based on the information currently available and the device inspection, it is most likely that the damage noted on the main coil during analysis and the reported event were due to the interaction with another coil inside the catheter during the clinical procedure.Therefore, an assignable cause of ¿operational context caused¿ will be assigned to the analyzed issue and reported event, as the issue is associated with a product that meets the design and manufacture specifications and was used in according with the dfu (direction for use) but due to procedural and/or anatomical factors during use, the device performance was limited.
|
| |
|
Event or Problem Description
|
|
During the coil embolization procedure, it was reported that the physician thought the first coil (subject device) was smaller than the size on the label.However, he still detached the coil.Then the physician advanced the second coil and big resistance was felt in the microcatheter.So he withdrew the second coil, but the first coil came out with the second coil together and the first coil was found stretched.The procedure was completed with other coils.No clinical consequences reported to the patient.
|
| |
|
Search Alerts/Recalls
|