| Model Number |
H749CBM3250100 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Device evaluated by mfr: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.A 10mmx2.50mm flextome® cutting balloon® was selected for use.During the procedure, it was noted that balloon ruptured at the second inflation.The procedure was completed with another device.No patient complications reported and patient's status was fine.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.A 10mmx2.50mm flextome® cutting balloon® was selected for use.During the procedure, it was noted that balloon ruptured at the second inflation.The procedure was completed with another device.No patient complications reported and patient's status was fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated: device avail.For eval, returned to mfr.On, device returned to mfr., device evaluated by mfr., eval summary attached, method codes, result codes and conclusion codes.Device evaluated by mfr: the device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Blood was identified within the balloon and lumen which is evidence of a device leak.The returned device was attached to an encore inflation unit and positive pressure was applied in an attempt to inflate the balloon.The balloon could not be inflated due to the presence of solidified blood that was present within the inflation lumen.The device was soaked in a water bath at a temperature of 37 degrees celsius to help soften the blood before further inflation attempts were made.The device was removed from the bath and the balloon was again attached to an inflation device subjected to positive pressure and liquid was observed to be leaking from a balloon pinhole located approximately 3mm proximal to the proximal edge of the distal markerband.An examination of the balloon material and markerband identified no issues which could potentially have contributed to this complaint.A visual and microscopic examination observed no damage to the tip or blades.All blades were present and fully bonded to the balloon surface.The hypotube was kinked at several locations along its length.This type of damage is consistent with excessive force being applied to the delivery system.No other issues were identified during the product analysis.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
| |
|