PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD Back to Search Results
Catalog Number 4002C3260
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure to treat a type i endoleak of aaa using penumbra coils 400 (pc400¿s).During the procedure, while attempting to advance the first pc400 through a non-penumbra microcatheter, the pc400 became stuck approximately one centimeter from the distal tip of the microcatheter; therefore, it was removed and inspected.The pc400 was then re-advanced into the microcatheter; however, the pc400 became stuck again at the same position.Therefore, the pc400 was removed and the procedure was completed using another pc400, six smart coils, a non-penumbra coil and the same microcatheter.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Results: the pet lock was intact on the proximal end of the pusher assembly.The embolization coil was intact with its pusher assembly and the introducer sheath had coagulated blood inside.Conclusions: evaluation of the returned pc400 revealed that the introducer sheath had coagulated blood inside.During the functional test the pc400 could not be advance through a demonstration microcatheter due to the coagulated blood inside the introducer sheath.Therefore, the root cause of the reported issue of the coil becoming stuck inside the microcatheter could not be determined.The non-penumbra microcatheter identified in the complaint was not returned to penumbra for evaluation.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA COIL 400
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7473574
Report Number3005168196-2018-00891
Device Sequence Number10791582
Product Code HCG
UDI-Device Identifier00814548012001
UDI-Public00814548012001
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/03/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date01/21/2025
Device Catalogue Number4002C3260
Device Lot NumberF73932
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 04/03/2018
Supplement Date Received by Manufacturer08/01/2018
Initial Report FDA Received Date04/30/2018
Supplement Report FDA Received Date08/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-