• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT RX HERCULINK ELITE STENT SYSTEM; RENAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT RX HERCULINK ELITE STENT SYSTEM; RENAL STENT SYSTEM Back to Search Results
Catalog Number 1011504-18
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: visual, dimensional and functional inspection was performed.The reported difficulties were confirmed.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other incidents.The investigation was unable to determine a conclusive cause for the reported difficulties.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported the procedure was performed to treat a lesion in the superficial mesenteric artery.The 7.0 x 18 mm herculink elite stent delivery system (sds) was advanced onto the guide wire; however, while attempting to advance into the 6f sheath, resistance was noted with the sheath prior to entering the anatomy.The sds was removed without issue and it was noted that a stent strut was flared up.The sds was no longer used and replaced.A new herculink elite sds was used without issue.There were no adverse patient effects and no clinically significant delay in the procedure.Returned device analysis identified that the stent was moved on the balloon.Returned device analysis identified that the stent was moved on the balloon.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX HERCULINK ELITE STENT SYSTEM
Common Device Name
RENAL STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7473688
Report Number2024168-2018-03158
Device Sequence Number1717837
Product Code NIN
UDI-Device Identifier08717648078354
UDI-Public08717648078354
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P110001
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 04/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2020
Device Catalogue Number1011504-18
Device Lot Number7100961
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/18/2018
Initial Report FDA Received Date04/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
Patient Weight77
-
-