• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. RUBY COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY4C0315
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance (2578)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/29/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the ruby coil pusher assembly was kinked and fractured along its length.Conclusions: evaluation of the ruby coils revealed an extensively damaged pusher assemblies.The embolization coil and non-penumbra catheter used during the procedure were not returned for evaluation.The root cause of the reported issue could not be determined due to extensive damage upon return.The majority of this damage was likely incidental and may have occurred during packaging for return to penumbra.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2018-00894.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the gastric and splenic artery using ruby coils.During the procedure, the physician successfully deployed and detached multiple ruby coils using a non-penumbra microcatheter.The physician then felt resistance and was unable to advance a ruby coil within the microcatheter, and therefore the physician removed the ruby coil, despite feeling resistance while retracting.This same issue then occurred with a second ruby coil.The procedure was completed using new ruby coils and the same microcatheter.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7473749
Report Number3005168196-2018-00893
Device Sequence Number18871990
Product Code HCG
UDI-Device Identifier00814548013152
UDI-Public00814548013152
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date10/30/2025
Device Catalogue NumberRBY4C0315
Device Lot NumberF80300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/03/2018
Initial Report FDA Received Date04/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
-
-