• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. EVERA AF MRI S; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. EVERA AF MRI S; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER Back to Search Results
Model Number DDMC3D1
Medical Device Problem Codes Inappropriate/Inadequate Shock/Stimulation (1574); Invalid Sensing (2293)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient had an elevated heart rate leading to t-wave oversensing (twos) and was shocked twice as a result.The lead remains in use.No patient complications have been reported as a result of this event.On (b)(6) 2018: it was further reported that about three months post implant of the patient¿s new implantable cardioverter defibrillator (icd) the patient received an inappropriate shock during an exercise period when hr (heart rate) was elevated and t-wave oversensing occurred.The icd did not withhold delivering shock therapy for the t-wave oversensing as the icd had determined it as a ventricular fibrillation (vf) episode.It was noted that all shock discriminators were on.The icd was reprogrammed and remains in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory indicated t-wave oversensing associated with the right ventricular lead.Analysis of the device memory indicated unexpected delivery of ventricular tachyarrhythmia therapy.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERA AF MRI S
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key7473817
Report Number3004209178-2018-09658
Device Sequence Number10802497
Product Code LWS
UDI-Device Identifier00643169717244
UDI-Public00643169717244
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P980016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/14/2018
Device Model NumberDDMC3D1
Device Catalogue NumberDDMC3D1
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/17/2018
Supplement Date Received by Manufacturer05/14/2018
Initial Report FDA Received Date04/30/2018
Supplement Report FDA Received Date05/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
694765 LEAD, 407652 LEAD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age26 YR
Patient Weight79
-
-