• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION ENVEO R DELIVERY SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION ENVEO R DELIVERY SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number ENVEOR-N-US
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Hematoma (1884)
Date of Event 04/09/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product analysis: the valve remains implanted and the delivery catheter system (dcs) was discarded by the customer; therefore, no product analysis can be performed.Conclusion: without return of the products, no definitive conclusions could be drawn regarding the clinical observation.Another relevant device is: product id: [evolutpro-26-us] serial/lot [(b)(4)] use by date [2019-08-31] udi [(b)(4)].If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that during the implant of this transcatheter bioprosthetic valve, when pulling the delivery catheter system (dcs) out from the valve after deployment, the nose cone became caught on the valve paddle and dislodged the valve.The valve was snared into the ascending aorta and a second transcatheter bioprosthetic valve was implanted.The patient developed a mediastinal hematoma that the physician reported was likely caused by snaring and pulling the initial valve to anchor it in the ascending aorta.No treatment was prescribed for the hematoma.No further adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENVEO R DELIVERY SYSTEM
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key7473873
Report Number2025587-2018-01020
Device Sequence Number16171521
Product Code NPT
UDI-Device Identifier00643169800939
UDI-Public00643169800939
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P130021
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/22/2018
Device Model NumberENVEOR-N-US
Device Catalogue NumberENVEOR-N-US
Device Lot Number0008892349
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2018
Initial Report FDA Received Date04/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age82 YR
-
-