• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC POLYAXIAL HEAD PLACEMENT TOOL FOR MATRIX; MISC ORTHO SURGICAL INSTR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC POLYAXIAL HEAD PLACEMENT TOOL FOR MATRIX; MISC ORTHO SURGICAL INSTR Back to Search Results
Catalog Number 03.632.037
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There was no known reported patient involvement associated with the complained event.It is unknown when device malfunctioned.Device is an instrument and is not implanted/explanted.The subject device has been received and the product evaluation is in progress.No conclusion can be drawn.A review of the device history records has been requested and is currently pending completion.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during setup for a procedure the scrub tech discovered that the matrix head application tool was broken.The internal piece on the distal working end had somehow snapped off and was rolling around in the tray.Fortunately, another instrument was available and was used in the procedure.There was no patient or procedure involvement, the malfunction was discovered prior to the procedure beginning.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Dhr review was completed part # 03.632.037 , synthes lot # 6603324 , supplier lot # 6603324 , release to warehouse date: 03oct2011 , supplier: (b)(4), no ncr's were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Customer quality conducted an investigation of the returned device.The polyaxial head placement tool (03.632.037) is used in the matrix degenerative and deformity spine systems.The tool is used to retrieve a polyaxial head from an implant module and place it over a matrix bone screw.The technique guides state that, to ensure the polyaxial head is securely attached to the bone screw, the user is to gently lift up on the placement tool and angulate the polyaxial head.In order to remove the instrument from the polyaxial head the button on the proximal end of the device can be pressed.The returned instrument was examined and the complaint condition was able to be confirmed as the blocker was found to be broken from the outer shaft and was returned.Further examination showed evidence of laser welding both the distal portion of the outer shaft and the blocker.The complaint condition was unable to be replicated due to post-manufacturing damage.No definitive root cause was able to be determined.The blocker is only laser welded to half of the outer shaft body, making it vulnerable to breakage in the area of the laser weld; as such the failure mode is consistent with rough handling.Relevant drawings for the returned instrument were reviewed: top-level, outer shaft w/ overmold and blocker.The design, materials and finishing processes were found to be appropriate for the intended use of these devices.The design, materials and finishing processes were found to be appropriate for the intended use of these devices.No dimensional analysis was possible due to post-manufacturing damage.A device history review, including material and hardness reviews, was performed for the returned instrument¿s lot number and no mrrs, ncrs or complaint-related issues were identified with the lot number which may have contributed to the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYAXIAL HEAD PLACEMENT TOOL FOR MATRIX
Common Device Name
MISC ORTHO SURGICAL INSTR
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7474011
Report Number2939274-2018-51956
Device Sequence Number1456486
Product Code LXH
UDI-Device Identifier10705034718958
UDI-Public(01)10705034718958
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.632.037
Device Lot Number6603324
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2018
Initial Date Received by Manufacturer 04/10/2018
Supplement Date Received by Manufacturer05/08/2018
Initial Report FDA Received Date04/30/2018
Supplement Report FDA Received Date05/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-