MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Medical Device Problem Code Air Leak (1008)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product event summary: the sheath was returned and analyzed.Visual inspection of the sheath showed that the device was intact with no apparent issues.Flushing/ air aspiration testing did not show any air passing through the tube or expelled from the distal tip.Functional testing of the sheath confirmed the hemostatic valve was leaking.Air bubbles were observed through the valve and it was suspected that the valve disk was torn.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during a cryo ablation procedure, during aspiration, air ingress was observed.The balloon catheter was prepped and inserted and air continued to be aspirated.The balloon catheter was maneuvered without resolve.The sheath was then replaced with resolve.The case was completed with cryo.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Common Device Name
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
MDR Report Key7474012
Report Number3002648230-2018-00274
Device Sequence Number13484775
Product Code DRA
UDI-Device Identifier00643169688872
UDI-Public00643169688872
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K123591
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/18/2020
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number42366
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/06/2018
Initial Report FDA Received Date04/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-