| Model Number |
186727855 |
| Medical Device Problem Code |
Disassembly (1168)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Screw head declamped (no breakage) intraoperatively from screw shaft during screw insertion (expedium cfx 186727855).Screw insertion was performed using the mis alignment device (2797-26-400) and an srt adapter and t handle.Screw head declamped half way in the pedicle.Screw shaft was then removed and a new screw was implanted using the regular screw driver.Patient consequence? unknown.Action taken for procedure:screw was removed and a new screw had to be implanted.Could have weakened the pedicle.New screw was augmented using confidence.Whether the screw will loosen must be monitored.Is the information being submitted for this complaint all the details that are known/available regarding this event? yes.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Visual examination of the returned device revealed that the screw head was disassembled from the shank, with broken saddle intact in the screw head.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.The root cause of the screw falling apart intraoperatively cannot be determined from the sample and the information provided.A potential root cause may be excessive force inadvertently placed on the drive feature of the screw during tightening, potentially at a significant angle, resulting in the screw disassembling during use.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|