STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number MICL13.2
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 04/04/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device evaluation: lens was returned in liquid, in lens vial.Visual inspection found a missing piece of haptic and a tear on a haptic.Patient code: (b)(4) secondary surgical intervention, explant and exchanged for a shorter length lens.(b)(4).
 
Event or Problem Description
The reporter indicated that a 13.2mm micl13.2, -11.0 diopter, implantable collamer lens was implaned into the patient's right eye (od) on (b)(6) 2018.The lens was explanted on (b)(6) 2018 due to over vault.The reporter stated that a shorter length lens was implanted, and the exchange resolved the problem.It was reported that there was no patient injury.At patient's last follow up uncorrected visual acuity was 20/20 and iop was normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7474236
Report Number2023826-2018-00655
Device Sequence Number18871380
Product Code MTA
UDI-Device Identifier00841542103312
UDI-Public00841542103312
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date04/30/2019
Device Model NumberMICL13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/04/2018
Initial Report FDA Received Date04/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age34 YR
-
-