Method: product history review, complaint history review, nc/capa history review, labelling review, risk assessment.Result: the reported event was confirmed based on the review of x-ray images.The reported device was not returned, therefore, product inspection could not be performed.Manufacturing records were reviewed for both lots and no relevant issues were identified.Conclusion: it cannot be determined when the cage fracture was initiated.The root cause of the reported cage fracture cannot be determined due to insufficient information and product not returned for analysis.Remedial action: stryker has decided to take a field action (3004024955-11/14/18-001-c).Specifically, stryker executed a correction by updating the surgical technique guide to warn end users against misuses which include extreme surgical techniques that fall outside the approved instructions for use.H3 other text : device not returned.
|