Model Number GTS100L
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 04/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(the stent was not implanted in the proper location and it is unknown where the stent is, therefore the surgery date was indicated in the implant date field.) the device remains in the patient, therefore product testing on the device could not be performed.The device history records were reviewed for this manufacturing lot and there were no non-conformities found to be related to the reported event.A part of the device history records review the sterilization records for this lot were reviewed and this device lot passed sterility test.In addition a complaint history review was completed for this lot, there were no complaints found to be related to the reported event.A review of the device labeling was completed.Malpositioned stents identified in the labeling as a known inherent risk of intraocular surgery.The ifu adequately provides instructions for stent implantation, precautions, and warnings for the proper use and handling of the device.Any omitted information was not available at the time of initial report.Written correspondence will be sent to the surgeon requesting the outstanding information.Subsequent attempts will be made as necessary (b)(4).
 
Event or Problem Description
As reported during a trabecular micro-bypass stent surgical procedure of the left eye, ¿there was too much blood¿ and the stent was lost either inside or outside of the patient¿s eye.The surgeon attempted to visualize the stent using an endoscopic camera, but was unable to locate the stent.The trabecular micro-bypass stent surgery was aborted.There was no patient injury.Additional information will be requested.
 
Additional Manufacturer Narrative
Mfr reference # (b)(4).
 
Event or Problem Description
Clinical follow-up was requested and on 05/02/2018 the surgeon provide the following clarification.Contrary to initial report of stent lost in the eye, it was learned that the stent was actually lost after removing it from the patient''s eye and therefore did not result in adverse impact to the patient.Since the stent was unable to be implanted in conjunction with cataract surgery, the surgeon performed endoscopic cyclophotocoagulation (ecp) as an alternative treatment for the patient¿s glaucoma.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISTENT TRABECULAR MICRO-BYPASS
Common Device Name
INTRAOCULAR PRESSURE LOWERING IMPLANT
Manufacturer (Section D)
GLAUKOS CORPORATION
229 avenida fabricante
san clemente CA 92672
MDR Report Key7474797
Report Number2032546-2018-00037
Device Sequence Number13716077
Product Code OGO
UDI-Device Identifier00853704002012
UDI-Public00853704002012
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P080030
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2020
Device Model NumberGTS100L
Device Catalogue NumberGTS100L
Device Lot Number113854
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/06/2018
Supplement Date Received by Manufacturer05/02/2018
Initial Report FDA Received Date04/30/2018
Supplement Report FDA Received Date05/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age75 YR
-
-