| Catalog Number |
257613100 |
| Medical Device Problem Codes |
Dull, Blunt (2407); Naturally Worn (2988)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Warringal private (b)(6) 2018 thread stripped -product code d2576134100 - 18x13 trial stem, unknown lot #.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The instrument associated with this report was not returned.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: product complaint # (b)(4).Investigation summary: the instrument associated with this report was not returned.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.A complaint was received regarding an srom dist stem trial 13 std stating: "thread stripped -product code d2576134100 - 18x13 trial stem, unknown lot #." update: the part was returned to the local engineering department in australia for evaluation: the end of the thread has been impacted causing the first thread to push backwards towards the second thread.This damage to the threads is preventing the neck trail from screwing on.This complaint is due to the impaction damaged caused to the thread.Photographs provided with the engineering assessment can be found in the pc-attachments.From the lot number on the device, mt29715, it was determined that the device was manufactured in october 2011 making it over 7 years old.No information received with this individual complaint indicated that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.The complaint shall be closed to wear and tear; entered into the complaints database and monitored through trend analysis.Post market surveillance is per sep 419.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|