| Model Number |
TRUETRACK |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-20-user's test strip had poor storage(kitchen).Test strip udi# (b)(4).Note manufacturer contacted customer on (b)(4) 2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.Dr.'s office staff member is calling on behalf of the customer.The customer is concerned with tests results from results obtained of 171 and 238 mg/dl.Dr.'s office staff member stated that the truetrack meter is also reading higher than their handheld meter which had read 123 mg/dl and the truetrack had read 179 mg/dl fasting.The customer's expected fasting blood glucose test result range is 100 - 120 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer.The product is not stored according to specification and is stored in the kitchen.The test strip lot manufacturer's expiration date is 06/21/2020 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (b)(6).
|
| |
|
Search Alerts/Recalls
|