(b)(4).Device was used for treatment, not diagnosis. the complaint devices were received and inspected.Visual observations revealed that a part of the needle is broken and stuck inside the shaft of the expressew iii and the needle flag is not attached with it.Hence this complaint can be confirmed.When the needle is jammed inside the device, it might have been pulled out from the distal end of the device which caused the needle to break.This operation is beyond the design intent of this device.The failure of this device can be attributed to the application error.A device history record review has been conducted for the expressew gun to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident and therefore there is no evidence of manufacturing anomalies on the paperwork reviewed.Further, a review into the depuy synthes mitek complaints system revealed one similar complaint and one dissimilar complaint for this lot of devices that were released to distribution.At this point in time, no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).Associated medwatch: 1221934-2018-50548.
|